Call Now 01423 781101 sales@fumecupboards-direct.com
Call us Monday - Friday: 8:30 am - 5:00 pm 01423 781101
Design - Manufacture - Install - Commission

K8 Technical Data

K8 Technical Data

Standard Widths1000mm1200mm1500mm1800mm2000mm
Sash Width700mm900mm1200mm1500mm1700mm
Sash Height500mm500mm500mm500mm500mm
Take Off Height2010mm2010mm2010mm2010mm2010mm
Take Off Diameter200mm200mm250mm250mm315mm
Centre from back200mm200mm200mm200mm200mm
Volume of air extracted0.175m3/sec

630m3/hour
0.175m3/sec

630m3/hour
0.175m3/sec

630m3/hour
0.175m3/sec

630m3/hour
0.175m3/sec

630m3/hour
Pa Loss6060606060
Weight200kg225kg250kg275kg300kg
External Dimensions2400mm(H)
x 900mm(D)
2400mm(H)
x 900mm(D)
2400mm(H)
x 900mm(D)
2400mm(H)
x 900mm(D)
2400mm(H)
x 900mm(D)
Internal DimensionsChamber height 1100mm
Working depth 720mm
Chamber height 1100mm
Working depth 720mm
Chamber height 1100mm
Working depth 720mm
Chamber height 1100mm
Working depth 720mm
Chamber height 1100mm
Working depth 720mm

Service connections

Services can be accessed by removal of the front service panels.

Broen Valves: Flexible hoses terminated at low level with 15mm copper

Vulcathene Bottle Trap: 38mm

Junction Box: 16A Isolated 230V electrical supply ( junction box located in top left hand side of fume cupboard )